The Bluelight s curing light is an example of the evolution in the technology used by Mectron Engineering.
The new light Bluelight s includes:
- The "Fast" function which allows high speed polymerisation.
- The "Slow Rise" function which allows a finely regulated power increase to alleviate problems of composite shrinkage.
- The possibility of Bleaching the teeth with products based on hydrogen peroxide.
These three functions have preset times and the possibility of storing, for each function, the most commonly used setting in the integrated memory.
Bluelight s is the only device available to incorporate a remote error detector RED.
This system produces sonar impulses which can transmit the diagnostic status of the unit, through a normal telephone line, to our after-sales technical centre.
TECHNICAL CHARACTERISTICS
Dimensions: 170 x 170 x 80 mm
Level of energy: (light intensity) > 800 mW/cm2
Spectral range: 400 - 510 nm peak intensity between 460 - 480 nm.
Fibre optic supplied: Ø 13-6,5 mm autoclavable (T.max 135°C).
Power supply: 230 Vac ± 10% 50/60 Hz - 66W - (optional 115 Vac - 100 Vac).
Halogen lamp: Philips type 13298 - 10 Volt - 52 W (minimum life 20 h).
Curing cycle: Continuous adjustment between 5" and 60" secs display. Acoustic signal at the end of the cycle. (10" - 60" in Slow Rise).
Mem button: to preset the most commonly used time for each function. Possibility of interrupting or repeating the cycle at any time.
Bleach: function possibility of bleaching the teeth with products based on hydrogen peroxide.
Forced cooling system: Length of cooling cycle proportional to the exposures.
Protections & Signals:
- Safety thermostat cut out with display indicator which automatically turns off the lamp and increases the cooling fan speed (E2 on display).
- E1 on display replace the (bulb).
- (E3-E4) on display over tension/under tension.
- (E7) Thermical protection.
Forced cooling: system length of cooling cycl
Diagnostic system: function R.E.D remote transmission of the operating parameters by sonar impulse.